1,0 /1,5 / 2,0 / 2,5 / 3,0 / 4,0mm
refill package for Test-equipment
Product information "Plates made of synthetic material"
Testing the elasticity - Ring-nosed forceps
Test: Place a piece of test material listed in table 2 between the functional surfaces in the anterior third of the working end. The ring-nosed forceps must be fully engaged (all ratchet teeth engaged) and stored at room temperature 3h +/- 5 minutes after being subjected to this stress. On completion of the test, there must not be any deformations or lasting changes noticeable.